Telemetry SNA and Pressure Transmitter (Small Animal, TR46SP)

Telemetry SNA and Pressure Transmitter (Small Animal, TR46SP)

Overview:

The MLE1022 Telemetry SNA and Pressure Transmitter (TR46SP) is suitable for measuring Sympathetic Nerve Activity and intravascular  pressure, in addition to temperature, in small animals ≥200 g (ideally suited to rats).

Compatible Receivers:

The MLE1022 transmitter is programmed with one of 12 independent transmission frequencies and is compatible with the following receivers:

Compatible Chargers:
The MLE1022 transmitter contains a rechargeable battery which can record continuously for approximately 6 hours. The transmitter battery is recharged within 4 h, or can be charged continuously for continuous data collection using the MLE1006CPS Wireless Continuous Power Supplier (Small Animal, TR802).

Compatible Telemetry Foundation Systems:

 

Features:

  • Fully implantable
  • A high gain amplifier, analog to digital converter (8 kHz sampling and an input range of ±60 micro Volts for the measured SNA signal, and 500 Hz sampling and an input range of -20 to +300 mmHg for the measured pressure signal)
  • An 8 cm fluid filled and gel tipped catheter
  • Wireless transmission range up to 5 m
  • Continuous charging/recording using the MLE1006CPS Wireless Continuous Power Supplier (Small Animal, TR802)
  • Transmitter power-down in situ, via the receiver, to conserve and extend battery life
  • Each transmitter is programmed with one of 12 independent transmitting frequencies
  • Transmitter battery is easily recharged using the MLE1006CPS Wireless Continuous Power Supplier, therefore no battery refurbishment is required.
This transmitter is capable of continuous charging/recording.

Applications:

The MLE1022 Telemetry SNA and Pressure Transmitter (TR46SP) is suitable for measuring Sympathetic Nerve Activity and intravascular or intralumenal pressure, in addition to temperature, in small animals ≥200 g (ideally suited to rats).

Specifications:


Input & Digitization
  • Input range: ±60 microV (SNA signal), -20 to +300 mmHg (pressure signal)
  • Resolution: 12 bit A/D resolution
  • Sampling: 8 kHz (SNA signal), 500 Hz (pressure signal)
  • Sensor: stainless steel (SNA electrodes), fluid filled with gel tipped catheter (pressure catheter)
  • Lead length: 15 cm (SNA electrodes), 8 cm (pressure catheter)

Battery Charge
  • Active mode: 6 hours
  • Standby mode: 6 months
  • Recharge time: ~4 h

Transmission
  • Frequency band: 2.4 GHz ISM band
  • Data latency: 20 ms
  • Range: up to 5 meters
  • Data integrity: >97%
  • Antenna: Omnidirectional

Physical
  • Weight: 13 g
  • Dimensions: 35 x 9 x 23 mm
  • Case material: medical grade silicone elastomer