Telemetry SNA and Pressure Transmitter (Small Animal, TR46SP)
- Product# MLE1022
- Add to Quote Request
Overview:
The MLE1022 Telemetry SNA and Pressure Transmitter (TR46SP) is suitable for measuring Sympathetic Nerve Activity and intravascular pressure, in addition to temperature, in small animals ≥200 g (ideally suited to rats).
Compatible Receivers:
The MLE1022 transmitter is programmed with one of 12 independent transmission frequencies and is compatible with the following receivers:
- MLE1162 SNA and Pressure Master Receiver Controller (TR162) and
- MLE1162SUB SNA and Pressure Subsidiary Receiver (TR162SUB)
Compatible Chargers:
The MLE1022 transmitter contains a rechargeable battery which can record continuously for approximately 6 hours. The transmitter battery is recharged within 4 h, or can be charged continuously for continuous data collection using the MLE1006CPS Wireless Continuous Power Supplier (Small Animal, TR802).
Compatible Telemetry Foundation Systems:
- PL3516B107 Telemetry SNA and Pressure Foundation System
Features:
- Fully implantable
- A high gain amplifier, analog to digital converter (8 kHz sampling and an input range of ±60 micro Volts for the measured SNA signal, and 500 Hz sampling and an input range of -20 to +300 mmHg for the measured pressure signal)
- An 8 cm fluid filled and gel tipped catheter
- Wireless transmission range up to 5 m
- Continuous charging/recording using the MLE1006CPS Wireless Continuous Power Supplier (Small Animal, TR802)
- Transmitter power-down in situ, via the receiver, to conserve and extend battery life
- Each transmitter is programmed with one of 12 independent transmitting frequencies
- Transmitter battery is easily recharged using the MLE1006CPS Wireless Continuous Power Supplier, therefore no battery refurbishment is required.
Applications:
The MLE1022 Telemetry SNA and Pressure Transmitter (TR46SP) is suitable for measuring Sympathetic Nerve Activity and intravascular or intralumenal pressure, in addition to temperature, in small animals ≥200 g (ideally suited to rats).Specifications:
Input & Digitization
- Input range: ±60 microV (SNA signal), -20 to +300 mmHg (pressure signal)
- Resolution: 12 bit A/D resolution
- Sampling: 8 kHz (SNA signal), 500 Hz (pressure signal)
- Sensor: stainless steel (SNA electrodes), fluid filled with gel tipped catheter (pressure catheter)
- Lead length: 15 cm (SNA electrodes), 8 cm (pressure catheter)
Battery Charge
- Active mode: 6 hours
- Standby mode: 6 months
- Recharge time: ~4 h
Transmission
- Frequency band: 2.4 GHz ISM band
- Data latency: 20 ms
- Range: up to 5 meters
- Data integrity: >97%
- Antenna: Omnidirectional
Physical
- Weight: 13 g
- Dimensions: 35 x 9 x 23 mm
- Case material: medical grade silicone elastomer