Thermistor Pod
- Product#
ML309
- Add to Quote Request
- Download Data Sheet
Overview:
The ML309 Thermistor Pod is suitable for biological measurements such as skin and nasal temperature and respiration detection in the range 5°C to 45°C. It also supports both absolute and relative temperature measurements. A DC offset provides the means for differential thermometry with resolutions of 0.001°C.
Compatible transducers include the:
MLT422/A Skin Temperature Probe (2 m)
MLT415/A Nasal Temperature Probe
MLT415/M Thermistor Temperature Sensor
The ML309 Thermistor Pod is not galvanically isolated.
Hardware:
ADInstruments signal conditioners only function with PowerLab data acquisition units and therefore are not suitable for use with other data acquisition systems.
Features:
- Voltage amplification to produce 50 mV/°C within the temperature range 5°C to 50°C.
- A 3-pole 10 Hz low-pass filter to reduce noise bandwidth.
- A switchable offset circuit which provides a delta temperature mode for relative temperature measurement.
Applications:
The Thermistor Pod is suitable for biological temperature measurements such as skin temperature measurements, nasal temperature and respiration detection.Specifications:
- Thermistor type: 10 kΩ @ 25 °C
- Measurement ranges: 5 °C to 45 °C in absolute mode Δ1 °C to Δ20 °C in five ranges with Delta Temp on
- Output voltage: 50 mV/°C (0 mV = 25 °C)
- Temperature accuracy: ±0.2 °C absolute (25 °C to 45 °C)
- DC drift: 3 μV/°C
- Response time: ~200 ms (for a full scale change in temperature)
- Overall response will be dependent on the thermal response of thermistor probe
- Temperature offset range: 50 °C overall (–5 °C to 45 °C absolute)
- Input connector: 3-pin mini audio connector
- Enclosure size: 108 x 58 x 35 mm (l x w x h)
- Weight: ~200 g
All specifications were tested at the time of printing and are subject to change.