Overview:
The FE136 Animal Bio Amp is a galvanically isolated, high performance, software-controlled differential amplifier suitable for the measurement of a wide variety of biological signals in animals and isolated tissues. Three shrouded 1.5 mm male pin sockets allow the direct connection of electrodes to this amplifier. Included with the FE136 Animal Bio Amp are MLA1212 Micro-Hook Electrodes for FE136 (3 pk).
The Animal Bio Amp can also be used with:
The MLAC14 Adapter Pack (2 mm Pin to SAF, 5 pk) is available (purchase separately) for connecting the FE136 Animal Bio Amp to any leads with 2 mm pins.
Hardware:
ADInstruments signal conditioners only function with PowerLab data acquisition units and therefore are not suitable for use with other data acquisition systems.
NOT FOR HUMAN CONNECTION or RECORDINGS
Features:
- A low-noise, high-gain differential amplifier specifically designed for biological signal measurements
- Software-controlled low-pass, high-pass and notch filters to remove unwanted signal frequencies for particular uses
- Audio output for use with EMG or EEG signals and similar
Applications:
Animal ECG, EMG and EEG, sensory nerve action potentials, visual evoked response, cortical evoked potentials and recording the long term electrical activity involved in involuntary muscle contractions as well as cardiac electrical activity in isolated heart experiments.Specifications:
Amplification
- Input amplifier: High impedance, differential (floating), electrically isolated
- Amplification range: ±5 μV to ±100 mV full scale in 14 steps
±100 mV 50 μV
±50 mV 25 μV
±20 mV 10 μV
±10 mV 5 μV
±5 mV 5 μV
±2 mV 1 μV
±1 mV 500 nV
±500 μV 250 nV
±200 μV 100 nV
±100 μV 50 nV
±50 μV 25 nV
±20 μV 10 nV
± 10 μV 5 nV
± 5 μV 2.5 nV
- Mid-band gain accuracy: ±1.5% (all ranges, within Bio Amp)
- Non-linearity: 0.1% within range
- Noise at various bandwidths:
0.3 Hz to 1 kHz: <0.6 μVrms
0.1 Hz to 100 Hz: <0.35 μVrms (@ 200 samples/second)
- CMRR: 85 dB (typical) @ 60 Hz
- IMRR: >130 dB (to true earth, 50 to 100 Hz)
Input
- Connection type: Three shrouded 1.5 mm male pin sockets
- Input impedance: 200 MΩ differential
- Input leakage current: <3 μArms @ 240 V, 50 Hz <2 μArms @ 120 V, 60 Hz
- DC blocking: ±1 V
- Baseline restore: Automatic
Filtering
- Low-pass filter: Fourth-order Bessel filter, ±3% accuracy
- Low-pass options: Software selectable. Standard: 50, 100, 200, and 500 Hz and 1 and 5 kHz (all at –3 dB); EEG mode: 3, 10, 30, 60, and 120 Hz
- High-pass filter: First-order filter, ±0.25% accuracy
- High-pass options: Software selectable. Standard: 0.1, 0.3, 1, 3, and 10 Hz (all at –3 dB); EEG mode: 0.03, 0.1, 0.3 and 1 seconds
- Notch filter: Second-order filter, –32 dB attenuation; 50 or 60 Hz frequency (automatic sensing)
Output
- Analog signal: ±2.0 V standard; ≤ ±4.0 V over range
- Audio: 3.5 mm (0.14") stereo jack; ±200 mV. Suitable for headphones or powered speakers
Control Port
- I2C port: Provides control and power. Interface communications rate of ~50 Kbits/s.
Physical Configuration
- Dimensions (h × w × d): 55 mm × 120 mm × 260 mm
- Weight: 1.2 kg
- Power requirements: <2 W
- Operating temperature range: 0 to 35 °C, 0 to 90% humidity (non-condensing)