NIBP System (for Rats)

NIBP System (for Rats)

Overview:

The ML125/R NIBP System is used in conjunction with a PowerLab system to obtain non-invasive blood pressure measurement from rats (with tail diameters 5-10 mm). It uses a specialized tail cuff and pulse transducer to intermittently measure blood pressure based on the periodic occlusion of tail blood flow.

The ML125/R system comes with a tail cuff and pulse transducer suitable for rats. The NIBP System outputs the pressure and pulse signals to a PowerLab via a BNC connection. LabChart software can be used to start the NIBP Controller at pre-set times. This is useful when the pulse is recorded continuously but blood pressure only needs to be sampled periodically.

The ML125/R NIBP System for rats includes a ML125 NIBP Controller and MLT125/R Pulse Transducer/Pressure Cuff for NIBP (Rat).
 
Suitable rat restrainers are available separately:

For a tail cuff system suitable for mice (with tail diameters 3- 6 mm), see the ML125/M page.

A MLA5030 Rodent Tail Cuff Holder, which allows the user to secure the rat tail transducer and cuff (MLT125/R) when attached to a rat tail, is also available separately.

Hardware:
The NIBP will only function with PowerLab data acquisition units and therefore is not suitable for use with other data acquisition systems.

Applications:

The NIBP System is used in conjunction with a PowerLab system to obtain non-invasive blood pressure measurement from rats (with tail diameters <10 mm).

This system operates by occluding blood flow to the tail using a specialized tail cuff and pulse transducer to intermittently measure blood pressure based on them periodic occlusion of tail blood flow.

For more information about this application see the Blood Pressure- Non Invasive page.

Specifications:

Cycle Operation
  • Max Inflation Pressure: 200 or 280 mmHg (switch-selectable)
  • Measurement Cycle Time:
40–150 BPM (large animals): 44 s (for 200 mmHg) 87 s (for 280 mmHg).
90–420 BPM (rats): 22 s (for 200 mmHg) 41 s (for 280 mmHg).
240–600 BPM (mice): 18 s (for 200 mmHg) 40 s (for 280 mmHg).
  • Cycle Control: Microprocessor based. Performs inflation, deflation, and fast deflation sequences automatically.
  • Operation Indication: Trigger output (normally low, but high during inflation and deflation cycle). Front panel Status indicator shows inflation and deflation operation.
  • Operation Abort: The cycle can be terminated at any point by pressing the front panel Start/Stop button again. Pressure is automatically released.
  • Fast Release Time:
~1.2 s from 280 to 40 mmHg at 40–150 BPM
~0.5 s from 280 to 40 mmHg at higher ranges.
  • Control Sources: Front panel push button; External signal source (voltage level); Remote contact closure.

Manual Start/Stop Input
  • Operation:Contact the closure input for starting or stopping measurement cycle.Shorting the input signal results in a start or stop operation.

Remote Trigger Input

  • Operation: Voltage level input for starting or stopping NIBP cycle. TTL compatible input. High level operates a start/stop.
  • Input Voltage: 3–5 V
  • Minimum Trigger Pulse: 1 ms

Trigger Output
  • Operation: High (+5V) output level during measurement cycle. Otherwise zero.

Pressure Output (Cuff Pressure)
  • Sensitivity: 0–1 V : 0–300 mmHg (factory calibrated)
  • Frequency Response: DC to 10 Hz

Pulse Input
  • Input Impedance: Differential Input: 10 GΩ
  • Single-ended Input: 1 MΩ
  • Input Signal Range: High gain differential input: (0–25 μV) up to (0–75 μV), depending on rear panel Gain Adj setting.
  • Low gain single-ended input: (0–50 mV) up to (0–150 mV), depending on rear panel Gain Adj setting.
  • Bandwidth:
40–150 BPM: 0.7–2.5 Hz
90–420 BPM: 3–7 Hz
240–600 BPM: 4–10 Hz
  • Pulse Output Max: ± 5 V

Operating Requirements
  • Power Requirements: PowerLab or MacLab I2C interface: +9 V @ 100 mA ±18 V @ 50 mA
  • Operating Conditions: 5–35 °C, 0–90% humidity (non condensing)

Physical Configuration
  • Dimensions (h × w × d): 65 mm × 200 mm × 275 mm (2.56" × 7.9" × 10.8")
  • Weight: 1.7 kg

ADInstruments reserves the right to alter these specifications at any time.