Oximeter Pod (Ear clip)
- Product#
ML320/E
- Add to Quote Request
- Download Data Sheet
Overview:
The ML320 Oximeter Pod is used for the non-invasive measurement of blood oxygen saturation (SpO2). The Oximeter Pod is a research device that is not intended for consumer use in the home.The Oximeter Pod plugs directly into a PowerLab Pod Port or the Pod Expander. Using the MLT322 SpO2 Ear Clip Sensor, direct saturation readings in the range of 70 to 100% can be made without the need to perform any other scaling.
This ML320/E is supplied with the MLT322 SpO2 sensor. For an Oximeter Pod with a Finger clip sensor please order the ML320/F.
For a unit configured for measuring pulse oximetry in animals the ML325/AW (tail wrap) or ML325/AC (Animal Clip) Oximeter Pods are available.
Hardware:
ADInstruments signal conditioners only function with PowerLab data acquisition units and therefore are not suitable for use with other data acquisition systems.
Features:
- Semiconductor light emitting diodes in the Ear-Clip transducer
- Pre-calibrated to read in % SpO2. The scale is fixed at 100% SpO2
- An alarm which will sound if the value of the SpO2 drops below 70% or a pulse could not be detected.
Applications:
The Oximeter Pod is suitable for the measurement of oxygen saturation in the range 70 to 100 %. For more information refer to the Pulse Oximetry application page.Specifications:
- Operating Principle: Non-invasive blood oxygen saturation (SpO2) determination using red and infrared light
- passed through pulsating blood in vascular tissue.
- Saturation range: 70 − 100%
- Accuracy (70 − 100% saturation): ±4% of full sclae using Ear Clip sensor
- Measurement wavelength: Red (660 nm) Infrared (910 nm)
- Measurement rate: 1 reading per second
- Output signal: 1.75 V for 98 %
- Resolution: 1 % steps
- Operating conditions: 0 – 35°C, 0 – 90 % humidity (non condensing)
- SpO2 acquisition time: ~10 second
All specifications were tested at the time of printing and are subject to change.