Oximeter Pod (Finger Clip)
- Product#
ML320/F
- Add to Quote Request
- Download Data Sheet
Overview:
The ML320/F Oximeter Pod (Finger Clip) is used for the non-invasive measurement of blood oxygen saturation (SpO2). The Oximeter Pod is a research device that is not intended for consumer use in the home.
The Oximeter Pod plugs directly into a PowerLab Pod Port or the ML305 Pod Expander. Using the MLT321 SpO2 Finger Clip, direct saturation readings in the range of 70 to 100% can be made without the need to perform any other scaling.
This ML320/F Oximeter Pod (Ear Clip) is supplied with the MLT321 SpO2 Finger Clip. For an Oximeter Pod with a finger clip sensor please order the ML320/E Oximeter Pod (Ear Clip) .
For a unit configured for measuring pulse oximetry in animals the ML325/AC Animal Oximeter Pod (Animal Clip), ML325/AW Animal Oximeter Pod (Wrap Sensor) or ML325/BS Animal Oximeter Pod (Base Sensor) are available.
Hardware:
ADInstruments signal conditioners only function with PowerLab data acquisition units and therefore are not suitable for use with other data acquisition systems.
Features:
- Semiconductor light emitting diodes in the Finger-Clip transducer
- Pre-calibrated to read in % SpO2. The scale is fixed at 100% SpO2
- An alarm which will sound if the value of the SpO2 drops below 70% or a pulse could not be detected.
Applications:
The Oximeter Pod is suitable for the measurement of oxygen saturation in the range 70 to 100 %. For more information refer to the Pulse Oximetry application page.Specifications:
- Operating Principle: Non-invasive blood oxygen saturation (SpO2) determination using red and infrared light
- passed through pulsating blood in vascular tissue.
- Saturation range: 70 − 100%
- Accuracy (70 − 100% saturation): ±2% for adults using Finger clip sensor
- Measurement wavelength: Red (660 nm) Infrared (910 nm)
- Measurement rate: 1 reading per second
- Output signal: 1.75 V for 98 %
- Resolution: 1 % steps
- Operating conditions: 0 – 35°C, 0 – 90 % humidity (non condensing)
- SpO2 acquisition time: ~10 second
All specifications were tested at the time of printing and are subject to change.