16 Channel Bio Amp

16 Channel Bio Amp

Overview:

The GT201/F16 Channel Bio Amp is a fully isolated, stand-alone biological amplifier that is optimized for the measurement of up to 16 biopotentials on a single subject. It consists of 8 modules, with each module containing two channel amplifiers that share a common ground input. The ground inputs from channels 1 to 8 are internally linked, as are channels 9 to 16. The hardware modules are provided with two ranges(500 µV/5 mV) and filter settings (Low-Pass: 100 Hz/1 kHz; High-Pass:0.5/2 Hz) that are selected via DIP-switches located at the rear of the unit.

LabChart (or Chart) software provides further adjustment of the amplifier range and filter settings to suit a variety of biological signals. This bioamplifier is particularly suitable for EEG, ECG/EKG, EMG and/or EOG recordings and can be used for both unipolar and bipolar recordings. It is supplied with a Sub-D to BNC cable for connecting each channel output to PowerLab data acquisition as well as a medical-grade power supply. An 8-Channel Bio Amp is also available.

Electrodes are NOT supplied with the 16-Channel Bio Amp and must be ordered separately. Suitable electrodes and lead wires include the:

Ordering Information:
When ordering please specify mains frequency F = 50 or 60 for 50 Hz or 60 Hz mains frequency.

For product specification and a summary of how to use the GT201 download the technique note. For a guide to the most appropriate bioamplifier configuration to suit your requirements download the bioamp selection guide.

Features:

  • Multi-channel amplifier for EEG, EMG, EOG and ECG recordings
  • Independently certified for connection to human subjects
  • 16 bipolar/unipolar channels
  • User-selectable filter and range settings via DIP-switches
  • Additional range and filter settings using Chart software
  • Direct analog-outputs suitable for connection to PowerLab data acquisition systems

Applications:

Any application in which the researcher wishes to record more than 2 biopotentials, such as EEG, EMG, EOG and ECG recordings from a single subject will require an 8 or 16 Channel Bio Amp. Common research applications include:

Sleep Studies
• Multichannel EEG
Psychophysiology

The 16 Channel Bio Amp is designed for use in teaching and research applications. It is NOT intended for clinical or critical life-care use and should NEVER be used for these purposes.

Specifications:

Input Safety sockets for biosignal electrodes

  • Output Output voltage: ± 5 V
  • Output connector: 25 pin Sub-D with 16 BNC connections (1.5 m cable)


Amplifier Module

  • 16 unipolar/bipolar channels
  • Noise level RTI: 0.7 μVss and 0.28 μVss for 0.1 to 10Hz
  • CMRR: > 110 dB
  • Subject Isolation: > 4 kV
  • Input Impedance: > 110 MOhm
  • Signal Amplification: μV up to mV range
  • Frequency Range: 0.5 Hz to 1 kHz


Settings

  • Selectable filter and range settings via DIP-switches
  • Signal Type- g.BSamp Setting:
Lo-Pass- 100 Hz/1 kHz
Hi-Pass- 0.5/2 Hz
Sensitivity- 500 μV/ 5 mV
  • 50 or 60Hz notch filter: on/off switchable


Physical Size

  • Length: 31 cm; Width: 25 cm; Height: 7 cm
  • Weight 2 kg
Products by Manufacturer:
Site Feedback
X

Site Feedback

http://www.adinstruments.com/products/hardware/research/product/GT201/

* Denotes a required field

* Name
* E-Mail
* Operating System
* Web Browser

* Comments