Single Channel Intracellular Preamplifier (IX1/115)
- Product# IX1
- Add to Quote Request
- Download Brochure
Overview:
The IX1 is a single-channel microelectrode preamplifier. It is suitable for a wide variety of applications including intracellular and extracellular recording. The preamplifier features a small-box headstage with a choice of five current ranges and a buzz with independent amplitude control to facilitate cell penetration. The optional Universal Buzz Box adds amplitude, duration and frequency features. A single intracellular headstage (7001/N) is included with the preamplifier. The preamplifier can also be used for extracellular recordings using the optional Extracellular Headstage. The preamplifier integrates seamlessly via the BNC analog outputs with PowerLab data acquisition systems. The unit is supplied with a wall mount power supply (115V AC only). This unit requires the PS-1 power supply for 220V operation (purchased separately).Please Specify: For the 7001 headstage select N = 0.001, 0.01, 0.1, 1 or 10
This equipment is not intended for use in human applications or human experimentation.
Features:
- Small Box Headstage with choice of five current ranges
- Buzz with independent amplitude control to facilitate cell penetration
- Small size takes up less space, yet is uncluttered, easy to use, with well defined controls
- Optional X1000 Voltage Gain AC Coupled Headstage for extracellular recording
Applications:
The Single Channel Intracellular Preamplifier is suitable for a wide variety of applications including intracellular and extracellular recordings.Specifications:
- Headstage size: 3.7 cm (1-1/2") X 2.9 cm (1-1/8") X 2.3 cm (7/8")
- Standard Headstage: N=.1 for 20 to 500 Megohm electrode. One Headstage included with IX1.
|
Current Multiplier N |
Max I |
RF (Feedback resistor ) |
RMS Noise / Bandwidth / Model cell Resistance |
Bridge Balance Range (ohms) |
Cp range |
Input Resistance |
|
10 |
10uA |
1Meg |
25uV,10Khz,1M |
10M |
20pF |
10Gig |
|
1 |
1uA |
10Meg |
50uV,10KHz,10M |
100M |
20pF |
100Gig |
|
.1 (standard) |
100nA |
100Meg |
50uV,1KHz,100M |
1G |
20pF |
1000 Gig |
|
.01 |
10nA |
1Gig |
40uV,100Hz,1G |
10G |
20pF |
10KGig |
|
.001 |
1nA |
10Gig |
35uV,10Hz,10G |
100G |
20pF |
100KGig |
- Filter Response: Switch selectable DC to 1kHz, 3kHz, 5kHz, 10kHz, 30kHz and Wide Band
- Input Bias Current: 1 picoamp (Adjustable to zero)
- Rise Time to Pulse: Typically 10 microseconds into 10 megohm resistor (10 to90%), 50 microseconds into 100 megohm
- Voltage Gain: X10
- Output Impedance: 500 ohms all outputs
- Offset Voltage: ±200mV
- Balance Reading: 0 to 100 megohm divided by headstage scaling N (10 turn digital dial)
- Electrode Impedance Test: 1mV per megohm square wave (N)
- Digital Polarization Current Range: 0 to +100 nA (10 turn digital dial) scaled by headstage multiplier N
- External Current Input BNC: 20nA/V scaled by the headstage multiplier
- Size: 7.3" wide, 2.6" high, and 9.7" deep.
- Power: Includes External Wall Mount Power Supply (115VAC 60Hz only). For 230VAC 50/60Hz, external power supply PS-1 is available.
- Electrode Holder: Low noise polycarbonate, HB180 for glass 1.5 to 1.8 mm O.D. (order separately)
- Extracellular Recording: Extracellular Headstage (AC coupled, voltage gain = 1000) (order separately)
- Optional: 7001 Spare Headstage (Specify N)