PowerLab 15T (LTS)

PowerLab 15T (LTS)

Overview:

The ML4818 PowerLab 15T is an integrated data recording unit featuring built-in amplifiers. With appropriate transducers and accessories the PowerLab 15T covers the broad experimental requirements in life science education.

The ML4818 PowerLab 15T includes:

 

 

 

Features:

  • Recording speed of up to 100 000 samples per second per channel (400 000 samples per second aggregate)
  • Two isolated biological inputs (approved for human connection, pending)
  • A constant-current isolated stimulator output (approved for human connection, pending)
  • Two non-isolated analog inputs
  • Two analog outputs
  • USB interface

Applications:

The PowerLab 15T can record biological signals from humans or animals, with full electrical isolation, and can also be used for any general-purpose stimulation. It is recommended for classroom applications such as Human and Animal Physiology, Neurobiology, Pharmacology, Psychophysiology, Biochemistry and more. For a comprehensive list of applications and experiments visit the PowerLab Education Applications page and LabChart Experiments page.

LabTutor Software and Teaching Experiments:

  • A revolution in teaching laboratory software, with data acquisition,experimental manual and report booklet in a single software interface.
  • Each LabTutor experiment includes background information, setup guides and student experimental protocols.
  • Real physiological data collection and experimentation performed using LabTutor panel.
  • Analysis and reporting sections are all contained within the LabTutor program
  • Provides a single-point management interface for educators on the number of courses, student data and reports via a single networked computer.


For more information visit the LabTutor page and the LabTutor Experiments page, for a list of current and the latest new and updated experiments.

Specifications:

(As tested at the time of printing and are subject to change)

Analog Inputs

  • Number of inputs: 2 DIN
  • Amplification:
Range
± 10 V
± 5 V
± 2 V
± 1 V
± 500 mV
± 200 mV
± 100 mV
± 50 mV
± 20 mV
Resolution
313 μV
156 μV
62.5 μV
31.3 μV
15.6 μV
6.25 μV
3.13 μV
1.56 μV
0.625 μV
  • Maximum input over-voltage: ± 15 V
  • Input impedance: 1 MOhm @ 150 pF (Inputs 1 & 2)
  • Maximum Bandwidth: 25 kHz
  • DC drift: Software corrected
  • CMRR: > 105 dB at a gain of 100
  • Interchannel crosstalk: > 90 dB
  • Signal-noise ratio: > 100 dB (10 V range)
  • Accuracy error: < 0.2%
  • Non-linearity (max) 0.1%
  • Filter – Low Pass: 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000 Hz and anti alias
  • Filter – Notch: Main


Pod Connectors

  • Inputs: 1 & 2
  • General features: Combined power, I2C and single ended or differential analog input signals on one connector, support SmartPod transducers.
  • Supply voltage: ± 5 V regulated
  • Maximum current: 50 mA per pod connector
  • Communications: 2-wire I2C
  • Signal input: Positive and negative analog inputs
  • Connector type: 8-pin DIN with metal surrounds


Bio Amp Input, Inputs 3 and 4

  • Number: 2
  • Input configuration: Differential channels with common isolated ground
  • Amplification:
Range
± 20 mV
± 10 mV
± 5 mV
± 2 mV
± 1 mV
± 500 μV
± 200 μV
± 100 μV
Resolution
625 nV
313 nV
156 nV
62.5 nV
31.3 nV
15.6 nV
6.25 nV
3.13 nV
  • Gain accuracy: 1%
  • Non-linearity: 1%
  • Bandwidth: 5 kHz
  • Noise: < 1 μVrms (0.5-2 kHz)
  • Maximum input over-voltage: ± 5 V
  • Input leakage current: < 4 μArms @ 240 V, 50 Hz < 4 μArms @ 120 V, 60 Hz
  • DC blocking: ± 0.3 V
  • Baseline restoration: Automatic or user-controlled
  • Input impedance: 100 MOhm to isolated ground (~500 pF per lead) using supplied Bio Amp subject cable and lead wires
  • Safety: IEC 60601-1 and CSA/UL approval pending
  • Isolation rating: 4000 V ACrms (1 minute)
  • IMRR: 130 dB
  • CMRR: 110 dB
  • High-pass filtering: Single-pole 0.5 Hz and 10 Hz
  • Low pass filtering: 10, 20, 50, 100, 200, 500, 1000, 2000 Hz


Sampling

  • ADC configuration: Each channel has its own ADC
  • ADC resolution: 24 bit
  • ADC linearity: ± 0.0006% FSR (INL)
  • Maximum sampling rates: 100 000 samples per second per channel (concurrent)– 400 000 samples per second (aggregate)
  • Available sampling rates: 100 000 samples per second down to 10 min/sample


Output Amplifier

  • Output configuration: 2 – outputs Complementary
  • Output resolution: 16-bit
  • Maximum output current: 20 mA
  • Accuracy error: < 0.2%
  • Linearity error: ± 0.5 LSB (INL)
  • Output ranges: ± 10 V
    • ± 5 V
    • ± 2 V
    • ± 1 V
    • ± 500 mV
    • ± 200 mV
  • Output slew rate: 2.3 V/μs
  • Settling time (G=1, 10 V step): 5 μs
  • Output impedance: 0.001 Ohms


Isolated Stimulator Output

  • Output configuration: Constant-current stimulator with hardware limited repetition rate
  • Isolation rating: 4000 Vrms to ground as per IEC 60601-1 2000 Vrms to Bio Amp inputs
  • Pulse duration: 50–200 μs (software-selectable)
  • Compliance voltage: 100–110 V typical
  • Output current: 0–20 mA in 0.1 mA steps (software adjustable)
  • Pulse rate: Software-selectable, but hardware-limited to a maximum of 20 Hz and 200 μs for safety
  • Safety indicators: Green and yellow indicators. Green flash indicates delivery of a valid stimulus. Concurrent green and yellow flash indicates an out-of-compliance condition.
  • Safety switch: Provides physical disconnection of the stimulator from the subject.


Microprocessor and Data Communication

  • CPU: Digital signal processor – Freescale DSP56858
  • RAM: 4 Mbit SRAM
  • ROM: 1 Mbit Flash ROM
  • Data communication: Hi-speed USB 2.0 (max 480 Mbit/sec transfer).
  • USB 1.1 compatible


Physical Configuration

  • Dimensions (w x h x d): 200 mm x 65 mm x 250 mm (7.9" x 2.6" x 9.8")
  • Weight: 1.8 kg (2.65 lbs)


Operating Requirements

  • Operating voltage range: 95 to 264 V, AC 47-63 Hz
  • Rated power: 25 VA
  • Operating conditions: 0 to 35 °C, 0 to 90% humidity (non-condensing).