PowerLab 15T (LTS)
- Product#
ML4818
- Add to Quote Request
- Download Data Sheet
- Download Product Manual
Overview:
The ML4818 PowerLab 15T is an integrated data recording unit featuring built-in amplifiers. With appropriate transducers and accessories the PowerLab 15T covers the broad experimental requirements in life science education.
The ML4818 PowerLab 15T includes:
- LabTutor Teaching Suite Software with LabTutor Experiments (one per server installation)
- Cable Kit including Power Cord, DIN to BNC Test Cable, USB Cable
- Getting Started with LabTutor Manual
- MLA2540 5-Lead Shielded Bio Amp Cable
- MLA2505 Shielded Lead Wires (5)
- MLADDF30 Stimulating Bar Electrode
- MLT1010/D Finger Pulse Transducer
Features:
- Recording speed of up to 100 000 samples per second per channel (400 000 samples per second aggregate)
- Two isolated biological inputs (approved for human connection, pending)
- A constant-current isolated stimulator output (approved for human connection, pending)
- Two non-isolated analog inputs
- Two analog outputs
- USB interface
Applications:
The PowerLab 15T can record biological signals from humans or animals, with full electrical isolation, and can also be used for any general-purpose stimulation. It is recommended for classroom applications such as Human and Animal Physiology, Neurobiology, Pharmacology, Psychophysiology, Biochemistry and more. For a comprehensive list of applications and experiments visit the PowerLab Education Applications page and LabChart Experiments page.
LabTutor Software and Teaching Experiments:
- A revolution in teaching laboratory software, with data acquisition,experimental manual and report booklet in a single software interface.
- Each LabTutor experiment includes background information, setup guides and student experimental protocols.
- Real physiological data collection and experimentation performed using LabTutor panel.
- Analysis and reporting sections are all contained within the LabTutor program
- Provides a single-point management interface for educators on the number of courses, student data and reports via a single networked computer.
For more information visit the LabTutor page and the LabTutor Experiments page, for a list of current and the latest new and updated experiments.
Specifications:
(As tested at the time of printing and are subject to change)
Analog Inputs
- Number of inputs: 2 DIN
- Amplification:
| Range ± 10 V ± 5 V ± 2 V ± 1 V ± 500 mV ± 200 mV ± 100 mV ± 50 mV ± 20 mV |
Resolution 313 μV 156 μV 62.5 μV 31.3 μV 15.6 μV 6.25 μV 3.13 μV 1.56 μV 0.625 μV |
- Maximum input over-voltage: ± 15 V
- Input impedance: 1 MOhm @ 150 pF (Inputs 1 & 2)
- Maximum Bandwidth: 25 kHz
- DC drift: Software corrected
- CMRR: > 105 dB at a gain of 100
- Interchannel crosstalk: > 90 dB
- Signal-noise ratio: > 100 dB (10 V range)
- Accuracy error: < 0.2%
- Non-linearity (max) 0.1%
- Filter – Low Pass: 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000 Hz and anti alias
- Filter – Notch: Main
Pod Connectors
- Inputs: 1 & 2
- General features: Combined power, I2C and single ended or differential analog input signals on one connector, support SmartPod transducers.
- Supply voltage: ± 5 V regulated
- Maximum current: 50 mA per pod connector
- Communications: 2-wire I2C
- Signal input: Positive and negative analog inputs
- Connector type: 8-pin DIN with metal surrounds
Bio Amp Input, Inputs 3 and 4
- Number: 2
- Input configuration: Differential channels with common isolated ground
- Amplification:
| Range ± 20 mV ± 10 mV ± 5 mV ± 2 mV ± 1 mV ± 500 μV ± 200 μV ± 100 μV |
Resolution 625 nV 313 nV 156 nV 62.5 nV 31.3 nV 15.6 nV 6.25 nV 3.13 nV |
- Gain accuracy: 1%
- Non-linearity: 1%
- Bandwidth: 5 kHz
- Noise: < 1 μVrms (0.5-2 kHz)
- Maximum input over-voltage: ± 5 V
- Input leakage current: < 4 μArms @ 240 V, 50 Hz < 4 μArms @ 120 V, 60 Hz
- DC blocking: ± 0.3 V
- Baseline restoration: Automatic or user-controlled
- Input impedance: 100 MOhm to isolated ground (~500 pF per lead) using supplied Bio Amp subject cable and lead wires
- Safety: IEC 60601-1 and CSA/UL approval pending
- Isolation rating: 4000 V ACrms (1 minute)
- IMRR: 130 dB
- CMRR: 110 dB
- High-pass filtering: Single-pole 0.5 Hz and 10 Hz
- Low pass filtering: 10, 20, 50, 100, 200, 500, 1000, 2000 Hz
Sampling
- ADC configuration: Each channel has its own ADC
- ADC resolution: 24 bit
- ADC linearity: ± 0.0006% FSR (INL)
- Maximum sampling rates: 100 000 samples per second per channel (concurrent)– 400 000 samples per second (aggregate)
- Available sampling rates: 100 000 samples per second down to 10 min/sample
Output Amplifier
- Output configuration: 2 – outputs Complementary
- Output resolution: 16-bit
- Maximum output current: 20 mA
- Accuracy error: < 0.2%
- Linearity error: ± 0.5 LSB (INL)
- Output ranges: ± 10 V
- ± 5 V
- ± 2 V
- ± 1 V
- ± 500 mV
- ± 200 mV
- Output slew rate: 2.3 V/μs
- Settling time (G=1, 10 V step): 5 μs
- Output impedance: 0.001 Ohms
Isolated Stimulator Output
- Output configuration: Constant-current stimulator with hardware limited repetition rate
- Isolation rating: 4000 Vrms to ground as per IEC 60601-1 2000 Vrms to Bio Amp inputs
- Pulse duration: 50–200 μs (software-selectable)
- Compliance voltage: 100–110 V typical
- Output current: 0–20 mA in 0.1 mA steps (software adjustable)
- Pulse rate: Software-selectable, but hardware-limited to a maximum of 20 Hz and 200 μs for safety
- Safety indicators: Green and yellow indicators. Green flash indicates delivery of a valid stimulus. Concurrent green and yellow flash indicates an out-of-compliance condition.
- Safety switch: Provides physical disconnection of the stimulator from the subject.
Microprocessor and Data Communication
- CPU: Digital signal processor – Freescale DSP56858
- RAM: 4 Mbit SRAM
- ROM: 1 Mbit Flash ROM
- Data communication: Hi-speed USB 2.0 (max 480 Mbit/sec transfer).
- USB 1.1 compatible
Physical Configuration
- Dimensions (w x h x d): 200 mm x 65 mm x 250 mm (7.9" x 2.6" x 9.8")
- Weight: 1.8 kg (2.65 lbs)
Operating Requirements
- Operating voltage range: 95 to 264 V, AC 47-63 Hz
- Rated power: 25 VA
- Operating conditions: 0 to 35 °C, 0 to 90% humidity (non-condensing).