Extracellular Recording System

Model No. ML870B73

Request price

The ML870B73 Extracellular Recording System allows recording of EEG, EKG and extracellular recording with the DP-311 Differential Amplifier with Active Headstage using glass or metal microelectrodes.

System Components:

DP-311 Features:
  • High input impedance
  • Excellent common mode rejection [120 dB @ 60 Hz]
  • High DC tolerance, high gain, low noise & bandwidth filtering
  • Small active headstage which can be mounted in micromanipulators
  • Suitable for use with high impedance glass or metal microelectrodes

High Input Z, High CMR:
High common mode rejection is very important in minimizing electrical interference. The high input resistance of the DP-311 Series (1012 Ohm typical) ensures that the instrument's high CMR will not be degraded by differences in source impedance at the input. This is important in extracellular recording, where the difference in resistance of the recording and indifferent electrode is often large.

High DC Tolerance:
Extracellular action potentials are typically measured in microvolts but are usually accompanied by much larger DC electrode voltages. The DP-311 is designed to amplify these microvolt level events with as much as ±2 V DC at the input.

High Gain, Low Noise & Bandwidth Limiting:
With gain selections at x10, x100, x1000 and x10000, even microvolt signals are sufficiently amplified for computer and recorder inputs. Noise is very low, typically 8 µV p-p at 1 Hz to 10 kHz bandwidth. Reducing the bandwidth with the low pass and high pass filters further lowers noise.

Calibration Signal:
An internal Cal signal provides a convenient check of amplifier gain and operation.


DP-311 Accessories:
  • 3110W Replacement Headstage for the DP-311 Differential Amplifier with Active Headstage
  • PJ2-5 2 mm Jack (pack of 5)


DP-311 Specifications:
Voltage Gain, AC & DC:                               x10, x100, x1000 & x10000
Input Resistance:                                        1012 Ω typical
Leakage Current:                                        1.0 pA typical
Common Mode Rejection:                            120 dB min (1000000:1) at 60 Hz
Noise, Input Shorted:                                  8 µV p-p, 1 Hz to 10 kHz typical
Low Frequency Filter Settings (high pass):     DC, 0.1, 1.0, 10 & 300 Hz
High Frequency Filter Settings (low pass):      0.1, 0.3, 1.0, 3.0, 10 & 50 kHz
Offset Control Range:                                  ±600 mV min at output, any gain set
Calibration Signal:                                       1 mV p-p at 100 Hz
Max. Common Mode Signal:                         ±2 V
Output Voltage Swing:                                 ±10 V (±5 V in gain of 10)
Output Resistance:                                      50 Ω
Input Connectors:                                       2 mm male pins, 0.5" (12.7 mm) spacing
Output Connector:                                       BNC
Power Requirements:                                  90 to 270 VAC, 50/60 Hz, 10 VA
Physical Size, H x W x D:                             5.1 x 21.25 x 18 cm
Shipping Weight:                                         2.3 kg
Warranty:                                                   Two years, parts and labor


Data Acquisition and Analysis:
The PowerLab 8/30 is shipped with ADInstruments software (LabChart and Scope), associated manuals and cables. The ML870/P is also supplied with LabChart Pro software which includes all the ADInstruments LabChart Modules (excluding GLP). LabChart Modules provide application specific data acquisition, analysis and display features.

The modules currently available include:

In purchasing this system containing LabChart Pro entitles the purchaser to 5 years of FREE LabChart Pro software upgrades and updates. Purchasers are also entitled to any ADInstruments modules released during this 5 year period (excludes GLP software).

For more information see the LabChart Pro page or contact your nearest ADInstruments representative.


Check out the ADInstruments Page on Facebook

ADResearch - Fast Tracking Research ADResearch by ADInstruments - Research Solutions for academic institutions, government organizations and private industry.
ADInstruments RSS News Feeds RSS  •  INTENDED USE  •  PRIVACY  •  LEGAL  •  SITE MAP  •  USER PROFILE  •  REGISTER