Bio Amps

Bio Amps

ECG / EKG, EEG, EOG, & EMG Biological Amplifiers

Galvanically isolated, high-performance differential biological amplifiers optimized for measuring a wide variety of biological signals such as ECG / EKG, EMG, EOG, and EEG, with high-quality noise specifications and a range of filter settings.

Overview

ADInstruments Bio Amps are galvanically isolated, high-performance differential biological potential amplifiers optimized to measure a wide variety of biological signals such as ECG / EKG, EMG, EOG, and EEG recordings.

In conjunction with our PowerLab data acquisition system options and LabChart analysis software, a wide range of filters are available for processing biosignals from a variety of research areas such as:

Our Single, Dual, Quad and Octal Bio Amps are supplied with Bio Amp Connection Cables and Lead Wires making them suitable and safe for both human and animal connection. An audio output is also available to help monitor signals.

Accessories suitable for use with Bio Amps include:

ADInstruments signal conditioners only function with a C Series Front End Interface or traditional PowerLab data acquisition device and therefore are not suitable for use with other data acquisition systems.

Amplifiers to Power Your Biological Data Acquisition

ADInstruments’ suite of Bio Amps are tailored to your acquisition and amplification needs. All amplifiers are galvanically isolated, and made specifically for the acquisition of bio signals such as action potentials and surface voltage fluctuation, powering the acquisition and recording of ECG, EMG, EOG and EEG. High input impedance and common mode rejection ratio (CMRR) support improved noise reduction over ultra-high bandwidth, increasing the ease of experiment setup and data acquisition.

Use your Bio Amp in conjunction with LabChart, a powerful software package that grants users ultimate customization in signal acquisition - from amplitude analysis to signal averaging.

Amplifiers connect to a set of data acquisition modules with PowerLab at the core. This easy integration of analog transducers and signal conditioners promotes a radically streamlined workflow.

Product Configurations

Bio Amp

Product Code
FE231
bftype
ce

The Bio Amp is a galvanically isolated, high-performance differential bio amplifier optimized for the measurement of a wide variety of biological signals such as ECG (EKG), EMG, EOG, and EEG recordings, with quality noise specifications and a range of filter settings.

$2,066.00
Ships from United States Estimated delivery 2-5 business days

Dual Bio Amp

Product Code
FE232
bftype
ce

The Dual Bio Amp consists of two galvanically isolated, high-performance differential amplifiers optimized for the measurement of a wide variety of biological signals such as ECG (EKG), EMG, EOG, and EEG recordings, with quality noise specifications and a range of filter settings.

$3,927.00
Ships from United States Estimated delivery 2-5 business days

Quad Bio Amp

Product Code
FE234
bftype
ce

The Quad Bio Amp consists of four galvanically isolated, high performance differential amplifiers optimized for the measurement of a wide variety of biological signals, with quality noise specifications and a range of filter settings.

$8,618.00
Ships from United States Estimated delivery 2-5 business days

Octal Bio Amp

Product Code
FE238
bftype
ce

The Octal Bio Amp consists of eight galvanically isolated, high performance differential amplifiers optimized for the measurement of a wide variety of biological signals, with quality noise specifications and a range of filter settings.

$14,529.00
Ships from United States Estimated delivery 2-5 business days

Bio Amp (with Front End Interface)

Product Code
FE231/CS
bftype
ce

The FE231/CS bundles the FE231 Bio Amp together with a C Series Front End Interface to make it compatible with PowerLab C.

Image

Dual Bio Amp (with Front End Interface)

Product Code
FE232/CS
bftype
ce

The FE232/CS bundles the FE232 Dual Bio Amp together with a C Series Front End Interface to make it compatible with PowerLab C.

Image

Quad Bio Amp (with Front End Interface)

Product Code
FE234/CS
bftype
ce

The FE234/CS bundles the FE234 Quad Bio Amp together with a C Series Front End Interface to make it compatible with PowerLab C.

Image

Octal Bio Amp (with Front End Interface)

Product Code
FE238/CS
bftype
ce

The FE238/CS bundles the FE238 Octal Bio Amp together with a C Series Front End Interface to make it compatible with PowerLab C.

Image

Key Features

  • Full electrical isolation from power-line (mains) circuitry to guarantee subject safety
  • A low-noise, high-gain differential amplifier specifically designed for biological signal measurements
  • Software-controlled low-pass, high-pass and notch filters to remove unwanted signal frequencies
  • Audio output for use with EMG or EEG signals and similar
  • Seamless integration with PowerLab data acquisition system options
  • Software Gain Control so your data is always correctly scaled in LabChart
  • Our Bio Amps are suitable for both human and animal connection
  • Compatible with LabChart 8.1.12 or above, and LabChart Lightning

Support

Support Articles

Our Support Articles contain answers to common questions, how-tos, solutions, and documentation.

Training and Workshops

Maximize time and resources with our customized training services delivered at your facility, on your equipment, on your terms.

See Training & Workshops

Details and Tech Specs

Technical

Quad Bio Amp Data Card

Octal Bio Amp Data Card

Input

  • Connection type: Three shrouded 1.5mm male pin sockets
  • Input configuration: 1 isolated differential channel with isolated ground reference or right leg drive
  • Input impedance: >1 GΩ differential, <100 pF (no cable) or 500 pF (supplied shielded Bio Amp cable and leads) to isolated ground
  • Isolation: 4000 Vrms (50 Hz for 1 minute)
  • Input ranges: ±100 μV to ±100 mV full scale in 10 steps (combined PowerLab and Bio Amp)
  • Gain accuracy: ±1.5 % all ranges
  • Non-linearity: <0.2% within range
  • Noise at various bandwidths:
    • 1 Hz to 10 kHz: <1.2 μVrms (<12 mV p-p)
    • 0.3 Hz to 1 kHz: <0.5 μVrms
    • 0.1 Hz to 100 Hz: <0.4 μVrms 
  • IMRR (isolation mode): >130 dB (to non-isolated earth, 50 Hz)
  • CMRR (common mode): >60 dB typical (150K electrode impedance, 5K imbalance @50Hz and 60Hz). >100dB typical (Balanced electrode impedance @50Hz and 60Hz)
  • Input leakage current: <4 μArms @ 240 V, 50 Hz 
  • DC tolerance: ±370 mV
  • Baseline restoration: Automatic or manual

Filtering

  • Low-pass filtering: Frequencies software-selectable: 50, 100, 200, 500, 1000, 2000, 5000, 10000 Hz, Off
  • High-pass filtering: Frequencies software-selectable: D.C., 0.0003, 0.001, 0.003, 0.01, 0.03, 0.1, 0.3, 1, 3,10, 30, 100, 200, 300 Hz
  • Adaptive mains filter: 50 or 60Hz frequency

Output

  • Signal: ±4.5 V maximum
  • Audio output: 3.5 mm (0.14") stereo output socket suitable for direct headphone or powered speaker connection. Output selectable from software. 

Control Port

  • I2C port:  Provides control and power. Interface communications rate of ~50 kbits/s.

Physical Configuration

  • Dimensions (h × w × d): 55 mm × 120 mm × 260 mm (2.2" × 4.7" × 10.2")
  • Weight: 1.3 kg (2 lb 12 oz)
  • Power requirements: ~2 W
  • Operating conditions: 5-35 °C. 0 to 90% humidity (non-condensing)
  • Transport / Storage conditions: 0-40 °C.  0-95% humidity (non-condensing)

Safety / Regulatory Information

  • Safety: Complies with IEC 60601-1:2012 
  • EMC: Complies with IEC 60601-1-2:2014 

Input

  • Connection type: Six-pin DIN/MS socket to fit 5-lead Bio Amp Cable (Tronomed D-1540)
  • Input configuration: 2 isolated differential channels with common isolated ground reference or right leg drive
  • Input impedance: >1 GΩ differential, <100 pF (no cable) to isolated ground or 500 pF(supplied shielded Bio Amp Cable and leads) to isolated ground
  • Isolation: 4000 Vrms (50 Hz for 1 minute)
  • Input ranges: ±100 μV to ±100 mV full scale in 10 steps (combined PowerLab and Bio Amp)
  • Gain accuracy: ±1.5 % all ranges
  • Non-linearity: <0.2% within range
  • Noise at various bandwidths:
    • 1 Hz to 10 kHz: <1.2 μVrms (<12 mV p-p)
    • 0.3 Hz to 1 kHz: <0.5 μVrms
    • 0.1 Hz to 100 Hz: <0.4 μVrms 
  • IMRR (isolation mode): >130 dB (to non-isolated earth, 50 Hz)
  • CMRR (common mode): >60 dB typical (150K electrode impedance, 5K imbalance @50Hz and 60Hz). >100dB typical (Balanced electrode impedance @50Hz and 60Hz)
  • Input leakage current: <4 μArms @ 240 V, 50 Hz 
  • DC tolerance: ±370 mV
  • Baseline restoration: Automatic or manual

Filtering

  • Low-pass filtering: Frequencies software-selectable: 50, 100, 200, 500, 1000, 2000, 5000, 10000 Hz, Off
  • High-pass filtering: Frequencies software-selectable: D.C., 0.0003, 0.001, 0.003, 0.01, 0.03, 0.1, 0.3, 1, 3,10, 30, 100, 200, 300 Hz
  • Adaptive mains filter: 50 or 60Hz frequency 

Output

  • Signal: ±4.5 V maximum
  • Audio output: 3.5 mm (0.14") stereo output socket suitable for direct headphone or powered speaker connection. Output selectable from software. 

Control Port

  • I2C port Provides control and power. Interface communications rate of ~50 Kbits/s.

Physical Configuration

  • Dimensions (h × w × d): 55 mm × 120 mm × 260 mm (2.2" × 4.7" × 10.2")
  • Weight: 1.3 kg (2 lb 12oz)
  • Power requirements: ~3 W
  • Operating conditions: 5-35 °C. 0 to 90% humidity (non-condensing)
  • Transport / Storage conditions: 0-40 °C.  0-95% humidity (non-condensing)

Safety / Regulatory Information

  • Safety: Complies with IEC 60601-1:2012 
  • EMC: Complies with IEC 60601-1-2:2014 

Input

  • Connection type: 9 x 1.5 mm (Quad) or 17 x 1.5 mm (Octal) single pin shrouded male socket
  • Input configuration: 4 (Quad) or 8 (Octal) isolated differential channels with common isolated ground reference or right leg drive
  • Input impedance: >1 GΩ differential, <100 pF (no cable) to isolated ground
  • Isolation: 4000 Vrms (50 Hz for 1 minute)
  • Amplification ranges: ±100 μV to ±100 mV full scale in 10 steps (combined PowerLab and Bio Amp)
  • Gain accuracy: ±1.5 % all ranges
  • Non-linearity: <0.2 % within range
  • Noise at various bandwidths:
    • 1 Hz to 10 kHz: <1.2 μVrms (<12 μV p-p)
    • 0.3 Hz to 1 kHz: <0.5 μVrms
    • 0.1 Hz to 100 Hz: <0.4 μVrms 
  • IMRR (isolation mode): >130 dB (to non-isolated earth, 50 Hz)
  • CMRR (common mode): >60 dB typical (150K electrode impedance, 5K imbalance @50Hz and 60Hz). >100dB typical (Balanced electrode impedance @50Hz and 60Hz)
  • Input leakage current: <4 μArms @ 240 V, 50 Hz 
  • DC tolerance: ±370 mV
  • Baseline restoration: Automatic or manual

Filtering

  • Low-pass filtering: Frequencies software-selectable: 50, 100, 200, 500, 1000, 2000, 5000, 10000 Hz, Off
  • High-pass filtering: Frequencies software-selectable: D.C., 0.0003, 0.001, 0.003, 0.01, 0.03, 0.1, 0.3, 1, 3,10, 30, 100, 200, 300 Hz
  • Adaptive mains filter: 50 or 60Hz frequency 

Output

  • Signal: ±4.5 V maximum
  • Audio output: 3.5mm stereo output socket suitable for direct headphone or powered speaker connection. Output selectable from software.

Control Port

  • I2C port: Provides control and power. Interface commun ications rate of ~50kbits/s.

Physical Configuration

  • Dimensions (h × w × d): 70 mm × 240 mm × 260 mm (2.7" × 9.45" × 10.2")
  • Weight: 2.5 kg (5.5 lb)
  • Power requirements: ~4.5 W (Quad), ~7 W (Octal)
  • Operating conditions: 5-35 °C. 0 to 90% humidity (non-condensing)
  • Transport / Storage conditions: 0-40 °C.  0-95% humidity (non-condensing)

Safety / Regulatory Information

  • Safety: Complies with IEC 60601-1:2012
  • EMC: Complies with IEC 60601-1-2:2014

ADInstruments reserves the right to alter these specifications at any time.

Product Codes

FE231
Bio Amp
FE232
Dual Bio Amp
FE136
Animal Bio Amp
FE234
Quad Bio Amp
FE238
Octal Bio Amp
FE231/CS
Bio Amp (with Front End Interface)
FE232/CS
Dual Bio Amp (with Front End Interface)
FE234/CS
Quad Bio Amp (with Front End Interface)
FE238/CS
Octal Bio Amp (with Front End Interface)

Compatibility

FE231

Operating System Recommended Hardware Alternate Hardware
LabChart Lightning Windows / Mac PowerLab C + Front End Interface Traditional PowerLab
LabChart 8 Windows / Mac PowerLab C + Front End Interface Traditional PowerLab

FE232

Operating System Recommended Hardware Alternate Hardware
LabChart Lightning Windows / Mac PowerLab C + Front End Interface Traditional PowerLab
LabChart 8 Windows / Mac PowerLab C + Front End Interface Traditional PowerLab

FE234

Operating System Recommended Hardware Alternate Hardware
LabChart Lightning Windows / Mac PowerLab C + Front End Interface Traditional PowerLab
LabChart 8 Windows / Mac PowerLab C + Front End Interface Traditional PowerLab

FE238

Operating System Recommended Hardware Alternate Hardware
LabChart Lightning Windows / Mac PowerLab C + Front End Interface Traditional PowerLab
LabChart 8 Windows / Mac PowerLab C + Front End Interface Traditional PowerLab