Human NIBP Controller

Get a quote

Thank you for your interest in Human NIBP Controller. Please complete and submit the form below and an ADInstruments representative will aim to get in touch within one working day.
If you’d like to speak to someone immediately, office contact information is available from the contact page.

×

Ask a question

Thank you for your interest in Human NIBP Controller. Please complete and submit the form below and an ADInstruments representative will aim to get in touch within one working day.
If you’d like to speak to someone immediately, office contact information is available from the contact page.

×

Book a demo

Thank you for your interest in Human NIBP Controller. Please complete and submit the form below and an ADInstruments representative will aim to get in touch within one working day.
If you’d like to speak to someone immediately, office contact information is available from the contact page.

×
×
checkmark Item successfully added to your quote request.

Human NIBP Controller

ML283

Human NIBP Controller

ML283

PRICE ON REQUEST

Manufactured by Finapres Medical Systems (FMS) and based on the Finapres 1000 Main Unit, the Human NIBP Controller is a high-end device that has been certified safe for human use.

Book a Demo

Overview

Manufactured by Finapres Medical Systems (FMS) and based on the Finapres 1000 Main Unit, the Human NIBP Controller is a high-end device that has been certified safe for human use.

The benefits of the Human NIBP Controller:

  • The Human NIBP Controller supports finger cuff switching between two Finger Cuffs, permitting lengthy recording durations with minimal discomfort to the subject.
  • With the Human NIBP Device Enabler installed, the Human NIBP Controller can stream data directly into LabChart. A PowerLab is not required. 
  • ADInstruments can supply the Human NIBP Controller and Human NIBP System (including LabChart Pro and Human NIBP Device Enabler) worldwide, whereas the Finometer MIDI has a limited distribution zone.
  • The Human NIBP System includes all the hardware and software to make continuous blood pressure measurements from human subjects, without any further purchases.

Details

Product Code


ML283
Human NIBP Controller

Technical


The Human NIBP Controller requires LabChart v8.0.4 or later

Physical Configuration

Dimensions (h × w × d): 58 × 191 × 249 mm  (2.3” × 7.5” × 9.8”)

Weight: 2600 grams

Power Requirements: 100 to 240 VAC, 47 to 65 Hz, 30W

Sample Rate: 200 Hz maximum

 

Environmental Specifications

Operating Temperature: 10–40 °C (50–104 °F)

Storage Temperature: -20–70 °C (-4–158 °F)

Humidity: 5–90% humidity (non condensing)

Ambient pressure: 700 to 1100 hPa

The device may not meet its performance specifications if stored or used outside the ranges specified above.

 

Instrumental Information

Cuff pressure: Max. 300 mmHg

Height sensing: Range ± 128 mmHg

Blood pressure accuracy: 1% of full scale (max. 3 mmHg);  Zeroing automatic

Finger cuff pump system: Pressure regulated at 380 mmHg;  Max. pressure: 420 mmHg

Analog output connector: BNC-type

 

Analog Output signals

Update frequency: 200 Hz

Time delay: < 5 ms

Offset: 0

Scaling: 100 mmHg/V

 

Analog Output connection

Analog output range: 0 to 4.0 V

Resolution: 2.5 mV

Offset: ± 5 mV

Internal impedance: < 1 Ω

 

Instrumental accuracy

Finger cuff pressure: 1% of full scale (max. 3 mmHg);  Automatic zeroing, typically, 0.5 mmHg

Height correction: 2% of full scale (max. 3 mmHg);  Manual zeroing

Heart rate: (Rate (bpm) /60%), i.e., at 60 bpm, accuracy is  ± 1%

Interbeat interval: 5ms (peak, non-accumulating)

 

Electrical Specifications

Power requirements: 100 – 240 VAC, 50/60 Hz, 50 VA

Main unit fuse: IEC 127, 0.8 A slow blow

Power cord: IEC 320 to local mains plug

Protection against electric shock (EN 60601-1):  Type of protection: Class I equipment

Degree of protection: type B, applied part

Power dissipation:

  • In Human NIBP Controller: < 50 W
  • In Wrist Unit: 1 W
  • In finger cuff: < 50 mW

 

Electromagnetic Compatibility

The Human NIBP Controller has been tested to comply with the requirements of EN60601-1-2 + A1: medical electrical equipment; EMC requirements and tests. The device also conforms to the following safety standards: CORR.2 (2007) and Canadian/US differences according to CAN/CSA-C22.2 No. 60601- 1:08, and ANSI/AAMI ES60601-1:2005.

The EMC conformity of the Human NIBP Controller includes the use of the following external cables, transducers and accessories (Non-observance may result in increased emissions or decreased immunity of the Human NIBP Controller):

Cable/accessory Description
Power cord Country specific type 
Wrist Unit cable assembly, including Height Correction Unit and finger cuff connected to the Wrist Unit

Standard length Wrist Unit cable assembly: 275 cm.

Standard length HCU cable: 30 cm.