Can I switch PowerLab channel inputs for ADInstruments Front-End Amplifiers with LabChart open?

The computer may have issues recognizing the Front-End Amplifiers/Signal Conditioners if you try to change channel inputs while a LabChart file is open. The correct way to change inputs is to do the following:

1) Save any necessary data and close LabChart. 
2) Turn off the PowerLab using the off/on switch on the back. 
3) Switch your amplifiers/Signal Conditioners to the desired PowerLab inputs. 
4) Turn the PowerLab back on. 
5) Open a new instance of LabChart, a settings file, or the file previously saved. LabChart should recognize the PowerLab and the new Front-End Amplifiers/Signal Conditioners input connection scheme upon opening. 

For further technical assistance with this or any other issue, please contact ADInstruments Technical Support by clicking HERE.