Firstly, check the wiring of the output cable. The center conductor must be the signal (+) and the outer conductor must be the ground/shield.
Measure the voltage directly at the BNC output connector using a voltmeter. The voltage outputs should be as follows:
Channel | Calibration Setting | Approximate Voltage |
Pressure 1 and 2 | 0 mmHg | 0 V |
Pressure 1 and 2 | 25 mmHg | 0.25 V |
Pressure 1 and 2 | 100 mmHg | 1 V |
ECG | 1 mV | 1 mV |
The Approximate Voltage and Calibration Setting for the Volume signal are largely dependent on the gain settings. The value selected in the Volume box on the Catheter Calibration screen in the Control Interface Software is the value that the MPVS Ultra is outputting. Verify that each signal is functional by observing the respective indicators on the front panel.