How do I use the MLA191 Calibration Kit to calibrate my new Laser Doppler Flow Probe?

New flow probes not initially purchased with an INL191 Laser Doppler Blood Flowmeter must be calibrated by following the procedure outlined below (Note: calibration instructions are included with the MLA191 Calibration Kit) with the flowmeter prior to the first use.

 
Initial Preparations
 
Before starting this it is advisable to have a stable workbench free from vibration. A clamp to hold the probe reliably in the suspension is provided in the calibration kit. 
 
Set up the probe in the center of the suspension, with the measuring surface of the probe at a maximum distance from all edges of the vessel. Note that calibration can be abandoned at any time by removing the probe or by pressing the CAL button on the flowmeter during the calibration.
 
Calibration
 
1. Ensure the probe has been properly initialized (left connected to the flowmeter) for at least 10 minutes prior to attempting the calibration (to allow for the temperature to stabilize)
 
2. Using the positioning clamps provided with the MLA191 Calibration Kitimmerse the probe tip to mid-depth in the calibration solution.
 
3. Press the CAL button once. A single beep is emitted, followed by a series of double beeps, lasting approximately 20 seconds.
 
4. While the beeping continues, press the CAL button once. A long beep will sound and the red and amber lights will flash for approximately 25 seconds.
 
5. A double beep will sound to indicate the end of the calibration, confirming the calibration. After five seconds, the red indicator light returns to amber and the probe is ready for use with this flowmeter.
 
During calibration keep the bench free of vibration and movement as this may invalidate the result. Generally, a probe does not require re-calibration unless a new or replacement probe is introduced.

For further technical assistance with this or any other issue, please contact ADInstruments Technical Support by clicking HERE.