How is the excitation of the ML221 Bridge Amp affected by the MLAC11 Grass Adapter?

The excitation voltage for a current generation ADInstruments Bridge Amplifier is set by an external resistor between Pin 5 and Pin 8 of the Bridge Amplifier's DIN connection.
 
While the Bridge Amplifier has a default excitation voltage of either 0V (Open circuit between setting pins) or 20V (short circuit), the MLAC11 Grass Adapter Cable has an internal resistor that will set this excitation properly for Grass Transducers.  To learn more about how to configure an ADInstruments Bridge Amplifier for use with non-ADInstruments transducers, please review Pages 29 - 34 of the Bridge Amplifiers Owner's Guide .

For further technical assistance with this or any other issue, please contact ADInstruments Technical Support by clicking HERE.