The Ventri-Cath line of Millar Catheters is not included in the default list of catheters and their parameters loaded when the Millar MPVS Ultra Interface software is installed. To use these catheters you will need to create a new entry for them under the Catheter Configuration Tab in the MPVS Ultra software. Listed below are the parameters for the Ventri-Cath 507/507S catheters:
New Configuration Name: User's Choice
Catheter Model: Ventri-Cath 507 or Ventri Cath 507S
Catheter Part Number: If 507 then "840-8221-07"; If 507S then "840-8221-07S"
Catheter Size: 5F
Catheter Length (cm): 120
Sensor Location: E4/E5
Electrode Spacing (mm): 7.0
Pigtail: If 507 enter "Yes"; If 507S enter "No"
Number of Electrodes: 12
Field Type: Dual
Excitation Current (uA): 100
Gain: x50