Why don't the force values in the DMT normalization module match those in the Chart View?

In some earlier versions of LabChart and the DMT normalization module, there was a problem if you used the prefix mN in the Units conversion when setting up LabChart for use with the DMT Wire Myograph systems.  Whilst the values would appear correctly in the Chart view with this conversion, when adding values to the DMT normalization module the value would be shifted by a factor of 1000.  

 
eg. 0.1mN in Chart view would appear as 0.0001mN in the normalization window.
 
Please make sure that you are using an up-to-date version of the DMT Normalization module; version 1.3.2 or later.
 
 