Why is the end systolic volume larger than the end diastolic volume in the PV-Loop Hemodynamic table?

The catheter is not fully in the left ventricle.  This temporally mis-aligns the pressure and volume data,and causes the PV-Loop module to incorrectly identify the EDV and ESV values.