Why is Fast Response Output only outputting one pulse when it is set for multiple pulses?

The potential cause for this issues is that too large of a hysteresis band is being calculated because the input range is set too high. The hysteresis value in the Fast Response Output extension is a percentage of the total channel range.  For example, if the input range is 10 V and the hysteresis is set at 10%, the hysteresis region is 1 V above or below the set activation level.  Change the range in LabChart to a smaller value that better suits the signal being input, and the Fast Response Output extension will output the pulses correctly.