Will the ML283 Human NIBP Controller / Cuff work with LabTutor 4?

The ML283 Human NIBP Controller can be used with LabTutor, if interfacing it to a PowerLab utilising the BNC Analogue Output of the Controller. The Analogue Output outputs the Raw Finger Pressure waveform, with a units conversion rate of 100mmHg / V. The units conversion, and general PowerLab setup can be configured using LabAuthor.

Note you will also need a copy of LabChart 8 running on the same machine in order for the Human NIBP Device Enabler to drive the functions of the NIBP unit.

Important note: In order for this configuration to work, you need to open LabTutor FIRST so that it has priority access to the PowerLab.