Cleaning and Sterilizing PowerLab Sensors

Clean PowerLab Sensors before and after each use to reduce the risk of contamination. Do not fully submerge PowerLab Sensors in liquids at any time.

General Cleaning

Use a lint-free cloth moistened with one of the following:

  • Industrial methylated spirit (denatured alcohol)
  • Disposable alcohol wipe
  • 70% ethanol spray (wipe off excess with a lint-free cloth)

Miscellaneous Cables

  • Wipe down using the general cleaning method above.

Velcro Straps

  • Rinse gently under warm water.
  • Allow to dry in moving air.
  • Do not submerge the sensor.

Biopotential Sensors, 12 Lead ECG Switch Box, GSR, and Lead Wires

  • Detach all biopotential and shielded lead wires
  • Clean using a lint-free cloth with alcohol or a disposable alcohol wipe
  • Avoid applying excess liquid to connector housings or transducers
  • Allow to dry completely before reconnecting

Transducers, Tendon Hammer, Stethoscope, Force Sensors, Goniometer, Push Button Switch, and Cardio Microphone

  • Clean the sensor body, cable, and connector housing using alcohol-based wipes or solution
  • Avoid contact with connection pins
  • Do not apply excessive liquid to connectors or transducers

Respiratory Belt Transducer

  • Hand wash the strap and cloth components with soap and water.
  • Do not immerse the electrical connector.
  • Avoid excess liquid near the connector housing.
  • Ensure the entire assembly is completely dry before use.

Spirometer Sensor

Refer to the dedicated cleaning protocol here.