What is the shelf-life of Kaha Sciences telemeters?

Rat Telemeters
Rat telemeters should be first implanted within 6 months of shipping from the factory, however it is best if they are used as soon as possible.

If a telemeter is to be stored for any length of time, ensure that it is first fully charged (battery level 90-100%). When storing telemeters, we recommend that the telemeter be turned on and the battery level checked regularly (at least every 2-3 months) using the Configurator System. After turning on the telemeter, check the battery level using the Configurator system and with the telemeter 1m away from the SmartPad. Before placing the telemeter back into Safe Mode, it should be fully charged (90-100%). After placing the telemeter into Safe Mode, make sure to keep it at least 1m away from any SmartPad to prevent it from turning back on.

Mouse Telemeters
Mouse telemeters can be stored long-term before their first use because there is no battery that can discharge over time.

For further technical assistance with this or any other issue, please contact ADInstruments Technical Support by clicking HERE.