Millar MPVS Duo: Hardware Set Up Demonstration

In this demonstration video Mark de Reus, Head of Training and Support at ADInstruments, shows you how to set up the MPVS Duo Hardware.

Step 1: LabChart Preparation

Step 2: Connecting the PowerLab C

  • Connect the C Series PowerLab and Instrument Interface to the PC. For more information on how to do this, watch the PowerLab C set up video.

Step 3a: Connecting Channels for Admittance Mode

  • Using the BNC cables, connect the following outputs from the MPVS Duo to the following inputs on the Instrument Interface:
    • Pressure to Input 1
    • Volume to Input 2
    • Phase to Input 3
    • Magnitude to Input 4
  • Connect your catheter to the MPVS Duo by inserting one side of the micro HDMI cable into the ‘catheter’ input on the MPVS Duo and the other into the catheter.
  • Plug the power supply into the MPVS Duo and then into the wall outlet.
  • Remember: Soak your catheter for at least half an hour before use.

Step 3b: Connecting Channels for Conductance Mode

  • Using the BNC cables, connect the following outputs from the MPVS Duo to the following inputs on the Instrument Interface:
    • Pressure to Input 1
    • Magnitude to Input 2
    • Phase to Input 3
  • Connect your catheter to the MPVS Duo by inserting one side of the micro HDMI cable into the ‘catheter’ input on the MPVS Duo and the other into the catheter.
  • Plug the power supply into the MPVS Duo and then into the wall outlet.
  • Remember: Soak your catheter for at least half an hour before use.

Step 4: Recording Data

  • The system is now ready to record PV data. Please see our other videos for more information on setting up the MPVS Duo and LabChart for recording.

Millar MPVS Duo: Admittance Set Up

Millar MPVS Duo: Conductance for Small Animals Set Up

Millar MPVS Duo: Conductance for Large Animals Set Up