Overview
A cost-effective, AC-coupled differential headstage for recording extracellular neuronal signals. It is ideally suited to teaching applications for advanced undergraduate neurophysiology.
Included with the Headstage:
- Stainless steel detachable electrode (1)
- Silver detachable electrode (1)
- Metal microelectrode male connectors (6)
Accessories purchased separately:
Key Features
- High impedance and a gain of 100x with a 100 Hz high pass filter and 5000 Hz low pass filter.
- Cable shielding that is directly connected to the casing, limiting the need for additional shielding at the input terminals.
- 3 input sockets (2 recording, 1 ground).
- Supplied with 6 metal microelectrode male connectors.
- Supplied with 2 pre-configured detachable electrode tips (1 stainless steel, 1 silver).
Related
Support
Support Assistant
Our Support Assistant can answer common questions and provide solutions.
Go to Support AssistantTraining and Workshops
Maximize time and resources with our customized training services delivered at your facility, on your equipment, on your terms.
See Training & WorkshopsDetails and Tech Specs
Technical
- Input: Differential
- Input connector: WPI 220-883-S02 to suit male 220-P02
- Impedance: 100 MΩ
- Gain: x100
- Low-pass filter: 5000 Hz (-3 dB frequency)
- High-pass filter: 100 Hz (-3 dB frequency)
- Input leakage current: ~1 pA
- CMRR: 97-100 dB, typical @ 200-1000 Hz
- Noise (rti): 100 Hz to 5 kHz <2 μVrms
- Power: from Port Input of PowerLab
- Electrode impedance: <300 k Ω, suitable for metal microelectrodes (purchased separately)
- Dimensions: Cylinder 8 mm outer diameter, 100 mm long
- Tube Material: Stainless Steel 304
- Electrode Tips: Stainless Steel 316, ~0.64 mm wire diameter (Black) Silver, ~0.64 mm wire diameter (Gray)
- Cable length: 2.9 m
Product Code
MLT185/ST