How do I reset my Masterflex L/S pump drive?

If your Masterflex Peristaltic Pump is not allowing you to correctly set the unit's pump direction or speed it may need to be reset.  To reset the Masterflex L/S pump drive, please follow the steps below:

 

  1. With the drive off, hold down the UP key and turn on the drive.
  2. The drive will alternately display “1” and then “CAL”,
  3. Release the UP key.
  4. Press the  UP or DOWN keys and then the display will scroll through the following maximum speed (in RPM) options:  100 -> 200 -> 300 -> 400 -> 600 -> 650.
  5. When the correct speed for your particular model is displayed, press the START/STOP key. To find the appropriate maximum speed for the model of Masterflex Peristaltic Pump you purchased from ADInstruments, please refer to the webpage located HERE.
  6. The display will momentarily light all segments and then return to zero.
  7. The maximum speed setting is now correct.

For further assistance with this issue, please reach out to your local ADInstruments Technical Support Representative.