Which pressure input on the CEC-10E PV Extension Cable should I use with my PV catheter?

The CEC-10E PV Extension Cable (10 ft) can connect a single-pressure or dual-pressure multi-segmented pressure-volume catheter, a single-pressure single-segmented pressure-volume catheter, and/or a pressure catheter to the MPVS Ultra PV Unit. It incorporates one REDEL connector for measuring the volume signal and two Low-Profile connectors for measuring one or two pressure signals.

If you are using a PV catheter with just one pressure sensor intended to measure ventricular pressure, you should connect it to the CEC-10E input labelled Pressure 1. If your PV catheter has a secondary pressure sensor, or if you are measuring pressure using a separate pressure catheter, you should connect the secondary pressure sensor's connector to the input labelled Pressure 2.