Can the IN191 / ML191 Laser Doppler Flow Meter & MN100XP Surface Probe be used on humans during exercise?

The IN191 Blood FlowMeter  and the MSP100XP Standard Surface Probe are suitable for use on stationary subjects only. The primary reason for this is that the probe is extremely susceptible to movement artefacts. Any minor movement is likely to cause large artefacts that will bury any actual signal. Securing the probe with adhesive rings / surgical tape may reduce the artefacts slightly, but this will likely still not be sufficient to reduce the artefacts to acceptable levels.

 
It is therefore recommended that the IN191 be used on stationary subjects only

For further technical assistance with this or any other issue, please contact ADInstruments Technical Support by clicking HERE.