The catheter's volume sensing segments are not fully in the left ventricle. This temporally mis-aligns the pressure and volume data, and causes the LabChart PV-Loop Module to incorrectly identify the EDV and ESV values.
For further technical assistance with this or any other issue, please contact ADInstruments Technical Support by clicking HERE.