Radnoti Micro Pressure Sensor

Radnoti Micro Pressure Sensor

Product Code 159905
The Radnoti Micro Pressure & Amplifier is designed to provide an alternative to the standard fluid-filled dome or flow through pressure transducers.
Book a Demo

Overview

The Radnoti Micro Pressure & Amplifier is designed to provide an alternative to the standard fluid-filled dome or flow through pressure transducers, especially in Isolated Rodent Liver/Kidney, Langendorff or Working Heart setups, where several pressure transducers must be used in close proximity in a temperature regulated environment.

Additionally, its compact, space-saving design incorporates a solid-state technology amplifier system. Hence, it has improved frequency response compared with the older style dome or flow through pressure sensors and amplifiers. Furthermore, the pressure transducer part is of a similar size to a Luer fitting that easily attaches into any female Luer port.

Therefore, it is ideal for addressing experiment issues of cost, ease of use and space, such as in isolated heart or blood vessel perfusion experiments.

Related

Research Applications

Langendorff
Cardiovascular Isolated Heart Systems
In Vitro Pharmacology + Isolated Tissue and Organs
Working Heart
Cardiovascular Isolated Heart Systems
In Vitro Pharmacology + Isolated Tissue and Organs
Isolated Organ
In Vitro Pharmacology + Isolated Tissue and Organs

Support

Support Articles

Our Support Articles contain answers to common questions, how-tos, solutions, and documentation.

Training and Workshops

Maximize time and resources with our customized training services delivered at your facility, on your equipment, on your terms.

See Training & Workshops

Details and Tech Specs

Technical

Specifications
Pressure range00-400 mmHg
Over pressure600 mmHg
Ambient pressure range670-800 mmHg
Analog input bandwidth frequency response drift1% of scale per 12 hours
Accuracy1% of Full Scale
Linearity0.1 mmHg
Analog bandwidth/sample rate pressure500Hz/1Khz; 2x over sampling
A/D Resolution20 Bits
Analog output0-3 VDC from 0-300 mmHg
Length of Cable16″ / 40 cm
Amplifier dimensions53x33x16 mm (LxHxW)
 

Product Code

159905
Micro Pressure Transducer & Amplifier

Requirements

Requires 159902-PWR power supply.

×
Ask a question

Have a question? Ask now

Ask a question